GB/T 13161-2015
Radiation protection instrumentation.Measurement of personal dose equivalents Hp(10) and Hp(0.07) for X, gamma,neutron and beta radiations.Direct reading personal dose equivalent meters (English Version)

Standard No.
GB/T 13161-2015
Language
Chinese, Available in English version
Release Date
2015
Published By
General Administration of Quality Supervision, Inspection and Quarantine of the People‘s Republic of China
Latest
GB/T 13161-2015
Replace
GB/T 13161-2003 GB/T 14323-1993
Scope
This standard applies to personal dose equivalent instruments with the following characteristics: a) worn on the human trunk or limbs; b) measuring the personal dose equivalent Hp(10 ) and Hp(0.07), and can measure personal dose equivalent rate?p(10) and?p(0.07); c) has a digital display; d) can have a personal Dose equivalent or personal dose equivalent rate alarm function. Therefore, this standard applies to the measurement of the following doses (including respective dose rates) and radiation combinations: 1) Hp(10) and Hp(0.07) produced by X and gamma radiation ; 2) Hp(10) and Hp(0.07) produced by X, γ and β radiation; 3) Hp(10) produced by X and γ radiation ; 4) Hp(10) produced by neutron radiation; 5) Hp(10) produced by X, γ and neutron radiation; 6) produced by X, γ and β radiation The resulting Hp(0.07). This standard specifies requirements for dosimeters and their associated readout systems, if provided. This standard specifies the general characteristics, general test methods, radiation characteristics and electrical, mechanical, safety and environmental characteristics of the above-mentioned dosimeters. Requirements affecting the accuracy of individual dose equivalent readouts and alarm settings and those involving the effect of the readout device on the dosimeter are specified only for the relevant readout system. This standard also stipulates the use classification of instruments according to different measurement capabilities in Appendix C. This standard does not include special requirements for accident or emergency dosimetry, although dosimeters may be used for this purpose. This standard does not apply to dosimeters for measuring pulsed radiation (for example: radiation produced by most medical X-ray diagnostic equipment, linear accelerators or similar equipment).

GB/T 13161-2015 Referenced Document

  • GB 4208-2008 Degrees of protection provided by enclosure(IP code)
  • GB/T 12162.1-2000 X and gamma reference radiation for calibrating dose-meters and doserate meters and for determining their response as a function of photon energy Part 1:Radiation characteristics and production methods
  • GB/T 12162.2-2004 X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy Part 2:Dosimetry for radiation protection over the energy ranges 8 keV to 1.3 MeV and 4 MeV to 9 MeV
  • GB/T 12162.3-2004 X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy Part 3:Calibration of area and personal dosemeters and the measurement of their response as a function of ener
  • GB/T 12162.4-2010 X and gamma reference radiation for calibrating dosemeters and doserate meters and for determining their response as a function of photon energy.Part 4:Calibration of area and personal dosemeters in low energy X reference radiation fields
  • GB/T 12164.1-2008 Reference beta radiations.Part 1:Methods of production
  • GB/T 14055.1-2008 Reference neutron radiations.Part 1:Characteristics and methods of production
  • GB/T 14055.2-2012 Reference neutron radiations.Part 2:Calibration fundamentals of radiation protection devices related to the basic quantities characterizing the radiation field
  • GB/T 16511-1996 Electrical and electronic measuring equipment--Documentation
  • GB/T 17626.11 Electromagnetic Compatibility Testing and Measurement Techniques Part 11: Voltage Dips, Short Interruptions and Voltage Variation Immunity Tests for Equipment with Input Current Less than or Equal to 16 A per Phase*2023-05-23 Update
  • GB/T 17626.4-2008 Electromagnetic compatibility.Testing and measurement techniques.Electrical fast transient/burst immunity test
  • GB/T 17626.5-2008 Electromagnetic compatibility.Testing and measurement techniques.Surge immunity test
  • GB/T 2900.81-2008 Electrotechnical terminology.Nuclear instrumentation Physical phenomena and basic conecpts
  • GB/T 2900.82-2008 Electrotechnical terminology.Nuclear instrumentation.Instruments, systems,equipment and detectors
  • GB/T 6592-2010 Electrical and electronic measurement equipment.Expression of performance

GB/T 13161-2015 history

  • 2015 GB/T 13161-2015 Radiation protection instrumentation.Measurement of personal dose equivalents H<下标p>(10) and H<下标p>(0.07) for X, gamma,neutron and beta radiations.Direct reading personal dose equivalent meters
  • 2003 GB/T 13161-2003 Direct reading personal dose equivalent(rate)monitors for X and gamma radiation
  • 1993 GB/T 14323-1993 Personal warning devices for x and gamma radiation
  • 0000 GB/T 13161-1991

GB/T 13161-2015 Radiation protection instrumentation.Measurement of personal dose equivalents H<下标p>(10) and H<下标p>(0.07) for X, gamma,neutron and beta radiations.Direct reading personal dose equivalent meters has been changed from GB/T 14323-1993 Personal warning devices for x and gamma radiation.

Radiation protection instrumentation.Measurement of personal dose equivalents H<下标p>(10) and H<下标p>(0.07) for X, gamma,neutron and beta radiations.Direct reading personal dose equivalent meters