CFR 40-1065.659-2014
Protection of Environment. Part1065:Engine-testing procedures. Section1065.659:Removed water correction.

Standard No.
CFR 40-1065.659-2014
Release Date
2014
Published By
US-CFR-file
Latest
CFR 40-1065.659-2014
Scope
(a) If you remove water upstream of a concentration measurement, x, correct for the removed water. Perform this correction based on the amount of water at the concentration measurement, xH2O[emission]meas, and at the flow meter, xH2Oexh, whose flow is used to determine the mass emission rate or total mass over a test interval. For continuous analyzers downstream of a sample dryer for transient and rampedmodal cycles, you must apply this correction on a continuous basis over the test interval, even if you use one of the options in § 1065.145(e)(2) that results in a constant value for xH2O[emission]meas because xH2Oexh varies over the test interval. For batch analyzers, determine the flow-weighted average based on the continuous xH2Oexh values determined as described in paragraph (c) of this section. For batch analyzers, you may determine the flow-weighted average xH2Oexh based on a single value of xH2Oexh determined as described in paragraphs (c)(2) and (3) of this section, using flowweighted average or batch concentration inputs.

CFR 40-1065.659-2014 history

  • 2014 CFR 40-1065.659-2014 Protection of Environment. Part1065:Engine-testing procedures. Section1065.659:Removed water correction.



Copyright ©2024 All Rights Reserved