This procedure is applicable to the initial calibration, subsequent calibration and in-use inspection of medical gamma-ray after-loaded brachytherapy radiation sources (dose rate range is approximately 0.01 Gy/min~5 Gy/min). This procedure does not apply to the verification of medical neutron short-range afterload therapy radiation sources, medical patch therapy radiation sources, and particle implant therapy radiation sources.
JJG 773-2013 Referenced Document
GB 4075 Sealed radioactive sources.General requirements and classification
GB 9706.13-2008 Medical electrical equipment.Part 2:Particular requirements for the safety of automatically-controlled brachytherapy afterloading equipment
GBZ 121-2002 Hygienic protection standard for after-loaded gamma source brachytherapy
JJF 1035 Metrological Terms and Their Definitions for Ionizing Radiation
JJG 773-2013 history
2013JJG 773-2013 Verification Regulation of Medical Radiation Source for γ-ray Afterloading Brachytherapy
1992JJG 773-1992 Verification Regulation of After Loading γ-Radiation Source for Brachytherapy