ASTM E2445-05(2010)
Standard Practice for Qualification and Long-Term Stability of Computed Radiology Systems

Standard No.
ASTM E2445-05(2010)
Release Date
2005
Published By
American Society for Testing and Materials (ASTM)
Status
Replace By
ASTM E2445/E2445M-14
Latest
ASTM E2445/E2445M-20
Scope

There are several factors affecting the quality of a CR image including the spatial resolution of the IP system, geometrical unsharpness, scatter and contrast sensitivity (signal/noise ratio). There are several additional factors (for example, scanning parameters), which affect the accurate reading of images on exposed IPs using an optical scanner.

The quality factors can be determined most accurately by the CR equipment manufacturer tests as described in Practice E2446. Individual test targets, which are recommended for practical user tests, are described for quality assurance. These tests can be carried out either separately or by the use of the CR phantom (Appendix X1). This CR phantom incorporates many of the basic quality assessment methods and those associated with the correct functioning of a CR system, including the scanner, for reading exposed plates and incorrectly erasing IPs for future use of each plate.

This practice is for users of industrial CR systems. This practice defines the tests to be performed, by users of CR systems, periodically to evaluate the CR systems to prove proper performance over the life-cycle of the system.

Application of Various Tests and Test Methods

Tests after Repair, Upgrade or the Use of Another IP Type:

Since modifications, such as repair or upgrade of the CR scanner and improved IP may improve the functionality of the system, specialized tests are required to prove the proper performance of the CR system.

User Tests for Long-term Stability8212;Quality assurance in test laboratories requires periodical tests of the CR system to prove the proper performance of the system. The time interval depends on the degree of usage of the system and shall be defined by the user and consideration of the CR equipment manufacturers information.

The tests described in 6.2.1 through 6.2.6 require usage of quality indicators of 5.1 or the CR test phantom shall be used regularly at user-defined intervals to test the basic performance. The documentation shall contain:

(1) Spatial resolution (by duplex-wire method, optional converging line pairs),

(2) Contrast (recognized contrast percentage of the material to examine),

(3) Slipping (yes/no),

(4) Jitter (yes/no),

(5) Shading (percentage at selected distance),

(6) Radiation parameters of the performed tests, and

(7) Date and operator name.

Fading tests should be performed only if the scanner or IP-brand is changed without data from the CR equipment manufacturer, or the system is used under extreme (beyond manufacturers recommendation) temperature conditions. The fading should be less than 50 % in the expected period between exposure and scan.

The IPs shall be checked for artifacts (6.2.7) and proper erasure (6.2.6).

Degradation of IPs or photo multipliers in the scanner may reduce the system sensitivity after extensive usage. For this reason, the SNR should be measured at longer intervals (for example, annual period) by the user or service personnel. The SNR shall not be less than 90 % of the original value. The increase of the SNR can be accepted without limits, if the system unsharpness is not increased.

1.1 This practice specifies the fundamental parameters of computed radiography systems to assure satisfactory and repeatable results for nondestructive testing.

1.2 This......

ASTM E2445-05(2010) history

  • 2020 ASTM E2445/E2445M-20 Standard Practice for Performance Evaluation and Long-Term Stability of Computed Radiography Systems
  • 2014 ASTM E2445/E2445M-14 Standard Practice for Performance Evaluation and Long-Term Stability of Computed Radiography Systems
  • 2005 ASTM E2445-05(2010) Standard Practice for Qualification and Long-Term Stability of Computed Radiology Systems
  • 2005 ASTM E2445-05 Standard Practice for Qualification and Long-Term Stability of Computed Radiology Systems