ASTM UOP703-09
Carbon on Catalysts by Induction Furnance Combustion and Infrared Detection

Standard No.
ASTM UOP703-09
Release Date
2009
Published By
American Society for Testing and Materials (ASTM)
Latest
ASTM UOP703-09
Scope

This method is for determining total carbon on catalysts and in other inorganic materials using a high temperature induction furnace carbon analyzer. The method is generally applied over a range of 0.01 to 20 mass-% carbon; however, up to 100 mass-% carbon can be determined with appropriate calibration. On used catalyst, the carbon may be present as coke and/or hydrocarbons. If only the carbon present as coke is desired, hydrocarbons are removed by extraction prior to analysis, as described in UOP Method 602, Soxhlet Extraction of Catalysts. Analysis of catalysts containing light hydrocarbons can be problematic due to thermal loss of the hydrocarbons before controlled combustion occurs. For such materials, other methods, such as ASTM D5291, Instrumental Determination of Carbon, Hydrogen, and Nitrogen in Petroleum Products and Lubricants, should be utilized.

Analyzers that utilize combustion tube, rather than induction furnace technology may be used if it can be proven that they yield precision data at least as good as that listed herein.

ASTM UOP703-09 history

  • 2009 ASTM UOP703-09 Carbon on Catalysts by Induction Furnance Combustion and Infrared Detection
  • 2002 ASTM UOP703-98 Carbon on Catalysts by Induction Furnace Combustion and Infrared Detection