YY/T 0695-2008
Standard test method for conducting cyclic potentiodynamic polarization measurements to determine the corrosion susceptibility of small lmplant devices (English Version)

Standard No.
YY/T 0695-2008
Language
Chinese, Available in English version
Release Date
2008
Published By
Professional Standard - Medicine
Latest
YY/T 0695-2008
Scope
This test method utilizes cyclic (forward and reverse) potentiodynamic polarization to evaluate the corrosion susceptibility of small metallic implantable medical devices or their components. Devices that can be evaluated by this test method include, but are not limited to, the following products: vascular stents, urethral stents, filters, endovascular graft support components, cardiac occluders, aneurysm clips and ligation clips, U-shaped staples, etc.

YY/T 0695-2008 Referenced Document

  • ASTM F1828 Standard Specification for Ureteral Stents
  • ASTM G15 Standard Terminology Relating to Corrosion and Corrosion Testing
  • ASTM G3 Standard Practice for Conventions Applicable to Electrochemical Measurements in Corrosion Testing
  • ASTM G5 Standard Reference Test Method for Making Potentiostatic and Potentiodynamic Anodic Polarization Measurements
  • ASTM G61 Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements for Localized Corrosion Susceptibility of Iron-, Nickel-, or Cobalt-Based Alloys*2024-04-19 Update
  • GB 6682 Analytical laboratory water specifications and test methods

YY/T 0695-2008 history

  • 2008 YY/T 0695-2008 Standard test method for conducting cyclic potentiodynamic polarization measurements to determine the corrosion susceptibility of small lmplant devices
Standard test method for conducting cyclic potentiodynamic polarization measurements to determine the corrosion susceptibility of small lmplant devices