ASTM E1329-00(2005)
Standard Practice for Verification and Use of Control Charts in Spectrochemical Analysis

Standard No.
ASTM E1329-00(2005)
Release Date
2000
Published By
American Society for Testing and Materials (ASTM)
Status
Replace By
ASTM E1329-10
Latest
ASTM E1329-10
Scope

Consistency in analysis depends on being aware of a significant change in instrumental response, such as that caused by drift or changes in analytical precision, or both, and taking corrective action. The usual corrective action for drift is standardization. Standardization, however, when there is no real need, can only broaden the spread of subsequent analyses. One purpose of this practice is to set guidelines that will avoid overstandardization.

To control manufacturing processes, there must be confidence that a consistent material is being produced and that the analysis of the material is reliable. For assurance that the material meets specification, a purchaser may require the supporting record of control charts to assess that proper analytical control has been maintained.

Ideally, variations in analytical results may be held to chance causes. The concept of a confidence interval or limits on a control chart is based on what can be expected when all normal precautions are exercised. When results appear to go out of control, the analyst should consider and correct what might be an assignable cause. As experience is accumulated, however, it may not seem unusual for readings to drift with time as optics degrade, detector response changes, or excitations modify as, for example, when deposits build up on a counter electrode (a correctable assignable cause), or the longer range effects as an X-ray tube deteriorates.

1.1 This practice covers procedures for determining if a spectrochemical analysis is under statistical control.

1.2 Criteria are presented for determining when corrective action is required.

1.3 Control will be effected by using verifiers to test instrument response. It is recommended, although not required, that this be accompanied by the plotting of control charts.

1.4 The preparation of control charts is described.

1.5 Limitations8212;The procedures that are described do not apply to analyses that require a calibration each time a set of analyses is run. Reference is made specifically to optical emission spectroscopy, but the practice has a more general application.

1.6 This practice does not apply to validation procedures that monitor the correctness of calibration.

ASTM E1329-00(2005) history

  • 2010 ASTM E1329-10 Standard Practice for Verification and Use of Control Charts in Spectrochemical Analysis
  • 2000 ASTM E1329-00(2005) Standard Practice for Verification and Use of Control Charts in Spectrochemical Analysis
  • 2000 ASTM E1329-00 Standard Practice for Verification and Use of Control Charts in Spectrochemical Analysis