ASTM D6091-07
Standard Practice for 99 %/95 % Interlaboratory Detection Estimate (IDE) for Analytical Methods with Negligible Calibration Error

Standard No.
ASTM D6091-07
Release Date
2007
Published By
American Society for Testing and Materials (ASTM)
Status
Replace By
ASTM D6091-07(2014)
Latest
ASTM D6091-07(2014)
Scope

Appropriate application of this practice should result in an IDE achievable by most laboratories properly using the test method studied. This IDE provides the basis for any prospective use of the test method by qualified laboratories for reliable detection of low-level concentrations of the same analyte as the one studied in this practice and same media (matrix).

The IDE values may be used to compare the detection power of different methods for analysis of the same analyte in the same matrix.

The IDE provides high probability (approximately 95 %) that result values of the method studied which exceed the IDE represent presence of analyte in the sample and high probability (approximately 99 %) that blank samples will not result in a detection.

The IDE procedure should be used to establish the interlaboratory detection capability for any application of a method where interlaboratory detection is important to data use. The intent of IDE is not to set reporting limits.

1.1 This practice establishes a standard for computing a 99 %/95 % Interlaboratory Detection Estimate (IDE) and provides guidance concerning the appropriate use and application. The calculations involved in this practice can be performed with DQCALC, Microsoft Excel-based software available from ASTM.

1.2 The IDE is computed to be the lowest concentration at which there is 90 % confidence that a single measurement from a laboratory selected from the population of qualified laboratories represented in an interlaboratory study will have a true detection probability of at least 95 % and a true nondetection probability of at least 99 % (when measuring a blank sample).

1.3 The fundamental assumption of the collaborative study is that the media tested, the concentrations tested, and the protocol followed in the study provide a representative and fair evaluation of the scope and applicability of the test method as written. When properly applied, the IDE procedure ensures that the 99 %/95 % IDE has the following properties:

1.3.1 Routinely Achievable IDE Value8212;Most laboratories are able to attain the IDE detection performance in routine analyses, using a standard measurement system, at reasonable cost. This property is needed for a detection limit to be practically feasible. Representative laboratories must be included in the data to calculate the IDE.

1.3.2 Routine Sources of Error Accounted for8212;The IDE should realistically include sources of bias and variation which are common to the measurement process. These sources include, but are not limited to: intrinsic instrument noise, some typical amount of carryover error, plus differences in laboratories, analysts, sample preparation, and instruments.

1.3.3 Avoidable Sources of Error Excluded8212;The IDE should realistically exclude avoidable sources of bias and variation, that is, those which can reasonably be avoided in routine field measurements. Avoidable sources would include, but are not limited to: modifications to the sample, measurement procedure, or measurement equipment of the validated method, and gross and easily discernible transcription errors (provided there was a way to detect and either correct or eliminate them).

1.3.4 Low Probability of False Detection8212;The IDE is a true concentration consistent with a measured concentration threshold (critical measured value) that will provide a high probability, 99 %, of true nondetection (a low probability of false detection, 945; = 1 %). Thus, when measuring a blank sample, the probability of not detecting the analyte would be 99 %. To be useful, this must be demonstrated for the particular matrix being used, and not just for reagent water.

1.3.5 Low Probability of False Nondetection8212;The IDE should be a true concentration at which there is a high probability, at least 95 %, of true detection (a......

ASTM D6091-07 history

  • 2007 ASTM D6091-07(2014) Standard Practice for 99?%/95?% Interlaboratory Detection Estimate 40;IDE41; for Analytical Methods with Negligible Calibration Error