Scope

This test method is useful for research and development, quality assurance, regulatory compliance and specification acceptance purposes.

The determination of the order of a chemical reaction or transformation at specific temperatures or time conditions is beyond the scope of this test method.

The activation energy results obtained by this method may be compared with those obtained from Test Method E 698 for nth order and autocatalytic reactions. Activation energy, pre-exponential factor and reaction order results by this method may be compared to those for Test Method E 2041 for nth order reactions.

1.1 Test Method A determines kinetic parameters for activation energy, pre-exponential factor and reaction order using differential scanning calorimetry from a series of isothermal experiments over a small (~ 10 K) temperature range. This treatment is applicable to low nth order reactions and to autocatalyzed reactions such as thermoset curing or pyrotechnic reactions and crystallization transformations in the temperature range from 300 to 900 K (30 to 630 176;C). This test method is applicable only to these types of exothermic reactions when the thermal curves do not exhibit shoulders, discontinuities or shifts in baseline.

1.2 Test Method B also determines the activation energy of a set of time-to-event and isothermal temperature data generated by this or other procedures.

1.3 Electronic instrumentation or automated data analysis systems or treatments equivalent to this test method may be used.

Note 1

Since all electronic data systems are not equivalent, the user must verify the applicability of the treatment to this method.

1.4 SI units are the standard.

1.5 This test method is similar but not equivalent to ISO 11357, Part 5, and provides more information than the ISO standard.

1.6 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use. Specific precautionary statements are given in Section 8.

ASTM E2070-03 history

  • 2023 ASTM E2070-23 Standard Test Methods for Kinetic Parameters by Differential Scanning Calorimetry Using Isothermal Methods
  • 2018 ASTM E2070-13(2018) Standard Test Method for Kinetic Parameters by Differential Scanning Calorimetry Using Isothermal Methods
  • 2013 ASTM E2070-13 Standard Test Method for Kinetic Parameters by Differential Scanning Calorimetry Using Isothermal Methods
  • 2008 ASTM E2070-08 Standard Test Method for Kinetic Parameters by Differential Scanning Calorimetry Using Isothermal Methods
  • 2003 ASTM E2070-03 Standard Test Method for Kinetic Parameters by Differential Scanning Calorimetry Using Isothermal Methods
  • 2000 ASTM E2070-00 Standard Test Method for Kinetic Parameters by Differential Scanning Calorimetry Using Isothermal Methods



Copyright ©2024 All Rights Reserved