Scope

1.1 The purpose of this specification is to ensure that reusable, detachable fiber-illuminated hook-on blade and handle combinations that engage shall lock and illuminate when in the operating position.

1.2 Dimensions and tolerances of fiber-illuminated systems shall be provided.

1.3 This specification provides methods of identifying by color the various fiber-illuminated systems and compatible blades and handles within the system.

1.4 This specification covers reusable handles and blades. It does not cover disposable components or systems.

1.5 Nonfiber-illuminated blades and handles with the lamp in the blade shall conform to Specification F965.

ASTM F1195-99 Referenced Document

  • ASTM F965 Standard Specification for Rigid Laryngoscopes for Tracheal Intubation-Hook-On Fittings for Laryngoscope Handles and Blades with Lamps