ASTM F2096-01e1
Standard Test Method for Detecting Gross Leaks in Medical Packaging by Internal Pressurization (Bubble Test)

Standard No.
ASTM F2096-01e1
Release Date
2001
Published By
American Society for Testing and Materials (ASTM)
Status
Replace By
ASTM F2096-02
Latest
ASTM F2096-11(2019)
Scope

1.1 This test method covers the detection of gross leaks in medical packaging. Method sensitivity is down to 250 m with an 81 % probability (see Section 11). This test method may be used for tray and pouch packages.

1.2 The sensitivity of this test method has not been evaluated for use with porous materials other than spunbonded polyolefin or with nonporous packaging.

1.3 This test method is destructive in that it requires entry into the package to supply an internal air pressure

1.4 The values stated in SI units are to be regarded as the standard. The values given in parentheses are for information only.

1.5

ASTM F2096-01e1 history

  • 2019 ASTM F2096-11(2019) Standard Test Method for Detecting Gross Leaks in Medical Packaging by Internal Pressurization (Bubble Test)
  • 2011 ASTM F2096-11 Standard Test Method for Detecting Gross Leaks in Packaging by Internal Pressurization (Bubble Test)
  • 2004 ASTM F2096-04 Standard Test Method for Detecting Gross Leaks in Medical Packaging by Internal Pressurization (Bubble Test)
  • 2002 ASTM F2096-02e1 Standard Test Method for Detecting Gross Leaks in Medical Packaging by Internal Pressurization (Bubble Test)
  • 2002 ASTM F2096-02 Standard Test Method for Detecting Gross Leaks in Medical Packaging by Internal Pressurization (Bubble Test)
  • 2001 ASTM F2096-01e1 Standard Test Method for Detecting Gross Leaks in Medical Packaging by Internal Pressurization (Bubble Test)
  • 2002 ASTM F2096-01 Standard Test Method for Detecting Gross Leaks in Medical Packaging by Internal Pressurization (Bubble Test)



Copyright ©2024 All Rights Reserved