These test methods provide information on the dimensional stability and wetting characteristics of autoclaved membrane filters. Excessive dimensional changes caused by autoclaving a constrained membrane filter could crack the filter, rendering it useless. Increased hydrophobicity, due to autoclaving would result in an increased time to filter a liquid sample.
1.1 These test methods deal with tests to determine the autoclavability of membrane filters.
1.2 These test methods are applicable to all membrane filters. Either Test Method A, Test Method B, or both, may be utilized to assess the autoclavability of membrane filters.
Section | |
Test Method A-Dimensional Stability | 8.1 |
Test Method B-Wetting Characteristics | 8.2 |
1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.