ISO 7206-10:2003
Implants for surgery - Partial and total hip-joint prostheses - Part 10: Determination of resistance to static load of modular femoral heads

Standard No.
ISO 7206-10:2003
Release Date
2003
Published By
International Organization for Standardization (ISO)
Status
Replace By
ISO 7206-10:2018
Latest
ISO 7206-10:2018/Amd 1:2021
Scope
This part of ISO 7206 applies to femoral heads of partial or total hip-joint replacements of modular construction (i.e. a head/neck conical taper connection) and describes methods of determining the load required, under specified laboratory conditions, to cause failure of the head (disassembly or fracture). It applies to components made of metallic and non-metallic materials. This part of ISO 7206 does not cover methods of examining and reporting the test specimens.

ISO 7206-10:2003 Referenced Document

  • ISO 4288:1996 Geometrical Product Specifications (GPS) - Surface texture: Profile method - Rules and procedures for the assessment of surface texture
  • ISO 6506-1:1999 Metallic materials - Brinell hardness test - Part 1: Test method
  • ISO 7206-1:1995 Implants for surgery - Partial and total hip joint prostheses - Part 1: Classification and designation of dimensions
  • ISO 7500-1:1999 Metallic materials - Verification of static uniaxial testing machines - Part 1: Tension/compression testing machines - Verification and calibration of the force-measuring system

ISO 7206-10:2003 history

  • 2021 ISO 7206-10:2018/Amd 1:2021 Implants for surgery - Partial and total hip-joint prostheses - Part 10: Determination of resistance to static load of modular femoral heads; Amendment 1
  • 2018 ISO 7206-10:2018 Implants for surgery - Partial and total hip-joint prostheses - Part 10: Determination of resistance to static load of modular femoral heads
  • 2003 ISO 7206-10:2003 Implants for surgery - Partial and total hip-joint prostheses - Part 10: Determination of resistance to static load of modular femoral heads
Implants for surgery - Partial and total hip-joint prostheses - Part 10: Determination of resistance to static load of modular femoral heads