This clause of Part 1 is applicable except as foIlows
1 1 Scope
Replacement:
This Part 2 is applicable to electrically powered LABORATORY CENTRIFUGES
1 1 1 AsDects excluded from the scope
—additional precautions which may need tO be obsewed when centrifuging materials which are
flammable or explosive(see 5 4 1 01):
一additional precautions which may need to be obsewed when centrifuging materials that
chemically react vigorously(see 5.4 101)
1 1 2 Equipmenf excluded from scope
—rotating electrical machinery within the scope of IEC 34
一CENTRIFUGES(used in explosive atmospheres)
1 2 Object
Addition:
一contact with moving parts fsee 7.2).
一CENTRIFUGE movement during any DISRUPTION(see 7 3 1 01).
一expelled parts(see 7 5):
一sinail particles from ROTOR ASSEMBLIES,expelled in air exhaust systems(see 7 5 104)
一high energy chemicaI reaction after ROTOR DISRUPTION(see 7 5 1 05).
一lneffectiveness of BIOSEALS(see 1 3 1 01)
1 4 Environmenta,conditions
Modifythethira dash asfollows:
一temperature+2。C to+40。C
CopyrighL by the UndenEiLers LaboraLories]nc
Sun Sep 05 15:36:10 2004
UL 3101-2-20-1997 history
1997UL 3101-2-20-1997 UL Standard for Safety Electrical Equipment for Laboratory Use; Part 2: Laboratory Centrifuges First Edition