GB/T 17437-1998 Procedures for calibrating and determining the response of neutron-measuring devices used for radiation protection purposes (English Version)
This standard specifies the methods for calibrating radiation protection neutron measuring instruments and determining the functional relationship between their response and incident neutron energy, angle and dose equivalent rate using the neutron reference radiation specified in GB 14055-93. According to the definition of ICUR 39 report and ICRU 43 report, neutrons of all energies are strong penetrating radiation, so the site monitoring instrument should be calibrated according to the surrounding dose equivalent Hp(10), while the personal dosimeter should be calibrated according to the penetrating personal radiation. Dose equivalent Hp (10) to calibrate. However, the method specified in this standard is general, and it can be used in various dose equivalent value systems. A schematic diagram of the physical properties to be tested is given in Appendix A (the standard's appendix).
GB/T 17437-1998 history
2012GB/T 14055.2-2012 Reference neutron radiations.Part 2:Calibration fundamentals of radiation protection devices related to the basic quantities characterizing the radiation field
1998GB/T 17437-1998 Procedures for calibrating and determining the response of neutron-measuring devices used for radiation protection purposes
GB/T 17437-1998 Procedures for calibrating and determining the response of neutron-measuring devices used for radiation protection purposes was changed to GB/T 14055.2-2012 Reference neutron radiations.Part 2:Calibration fundamentals of radiation protection devices related to the basic quantities characterizing the radiation field.