Scope
1 This International Standard specifies the performance requirements of RADIOTHERAPY DOSIMETERS, as defined in 3.1, intended for the measurement of ABSORBED DOSE TO WATER or AIR KERMA (and their RATES) in photon or electron radiation fields as used in radiotherapy. NOTE Throughout this standard: —if no material is specified, the term "ABSORBED DOSE" or "DOSE" means "ABSORBED DOSE TO WATER (in water)" and the term "KERMA" means "AIR KERMA (in air)"; —when the quantity "AIR KERMA (in air)" in units "Gy" is used, the quantity "EXPOSURE" in units "C/kg" is also allowable. 2 The dose-monitoring systems incorporated in radiotherapy treatment machines are not covered by this standard, neither are the re-entrant ion chambers used for brachytherapy source calibration. 3 This standard is applicable to the following types of DOSIMETER: a) FIELD-CLASS DOSIMETERS normally used for: 1) the measurement of KERMA or DOSE in a radiation beam, either in air or in a PHANTOM; 2) in vivo skin surface or intracavitary measurements of DOSE on PATIENTS; b) REFERENCE-CLASS DOSIMETERS normally used for the calibration of FIELD-CLASS DOSIMETERS. NOTE REFERENCE-CLASS DOSIMETERS may be used as FIELD-CLASS DOSIMETERS.

BS EN 60731:1997 history

  • 2022 BS EN 60731:2012+A1:2022 Medical electrical equipment. Dosimeters with ionization chambers as used in radiotherapy
  • 2012 BS EN 60731:2012 Medical electrical equipment. Dosimeters with ionization chambers as used in radiotherapy
  • 1997 BS EN 60731:1997 Medical electrical equipment. Dosimeters with ionization chambers as used in radiotherapy
Medical electrical equipment. Dosimeters with ionization chambers as used in radiotherapy



Copyright ©2024 All Rights Reserved