This standard specifies in detail the instruments and test procedures used for the absolute calibration of the acceleration sensor by the reciprocity method. This standard applies to linear acceleration sensors with a calibration frequency range of 40 to 1250 Hz and an acceleration of 10 to 100 m/s 2. The calibration uncertainty of this method is 1% when the reference point (160Hz, 100m/s2) and the amplifier gain gear are at the reference position.
GB/T 13823.18-1997 history
2008GB/T 20485.12-2008 Method for the calibration of vibration and shock transducers.Part 12: Primary vibration calibration by the reciprocity method
1997GB/T 13823.18-1997 Methods for the calibration of vibration and shock pick-ups. Primary vibration calibration by reciprocity method
GB/T 13823.18-1997 Methods for the calibration of vibration and shock pick-ups. Primary vibration calibration by reciprocity method was changed to GB/T 20485.12-2008 Method for the calibration of vibration and shock transducers.Part 12: Primary vibration calibration by the reciprocity method.