This part of ISO 7206 describes a method of deter-mining the torque required, under specified laboratory conditions, to loosen the fixation of the head of hip joint prostheses in which the head is not intended to rotate relative to the neck. It applies to the femoral component of total or partial hip joint replacements in which the head and arm are secured together by a locking conical taper or any other means and in which the head and neck are separate components, and which are made of metallic and non-metallic materials.
This part of ISO 7206 does not cover methods of ex-amining the test specimens; these should be agreed between the test laboratory and the party submitting the specimen for test.
BS 7251-13:1995 history
1995BS 7251-13:1995 Orthopaedic joint prostheses - Method for determination of resistance to torque of head fixation of stemmed femoral components