ASTM F1841-19e1
Standard operating procedure for assessment of hemolysis in continuous flow blood pumps

Standard No.
ASTM F1841-19e1
Release Date
2019
Published By
American Society for Testing and Materials (ASTM)
Latest
ASTM F1841-19e1
Scope
1.1 This practice covers a protocol for the assessment of the hemolytic properties of continuous, intermittent, and pulsatile flow blood pumps used in circulatory assist, including extracorporeal, percutaneous, and implantable devices. An assessment is made based on the pump’s effects on the erythrocytes over a certain period of time. Adopting current practices for this assessment, a 6-hour in vitro test is performed on a pump placed in a device-specific recirculating blood loop that mimics the pressure and flow conditions of the expected worst-case clinical use of the device. If the ultimate goal of the testing is to evaluate the blood damage potential of a pump for clinical use, it is suggested that paired testing between the subject blood pump and a legally marketed comparator device be conducted using the same blood pool in a matched blood test loop so that a relative hemolysis comparison can be made. 1.2 The values stated in either SI units or inch-pound units are to be regarded separately as standard. The values stated in each system may not be exact equivalents; therefore, each system shall be used independently of the other. Combining values from the two systems may result in non-conformance with the standard. 1.3 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.4 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the 1 This practice is under the jurisdiction of ASTM Committee F04 on Medical and Surgical Materials and Devices and is the direct responsibility of Subcommittee F04.30 on Cardiovascular Standards. Current edition approved Sept. 1, 2019. Published December 2019. Originally approved in 1997. Last previous edition approved in 2017 as F1841 – 97 (2017). DOI: 10.1520/F1841-19E01. Copyright © ASTM International, 100 Barr Harbor Drive, PO Box C700, West Conshohocken, PA 19428-2959. United States This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee. 1 Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

ASTM F1841-19e1 Referenced Document

  • ASTM F1830 Standard Practice for Selection of Blood for in vitro Evaluation of Blood Pumps
  • ISO 14708-5 Implants for surgery — Active implantable medical devices — Part 5: Circulatory support devices

ASTM F1841-19e1 history

  • 2019 ASTM F1841-19e1 Standard operating procedure for assessment of hemolysis in continuous flow blood pumps
  • 2019 ASTM F1841-19 Standard Practice for Assessment of Hemolysis in Continuous Flow Blood Pumps
  • 2017 ASTM F1841-97(2017) Standard Practice for Assessment of Hemolysis in Continuous Flow Blood Pumps
  • 1997 ASTM F1841-97(2013) Standard Practice for Assessment of Hemolysis in Continuous Flow Blood Pumps
  • 1997 ASTM F1841-97(2005) Standard Practice for Assessment of Hemolysis in Continuous Flow Blood Pumps
  • 1997 ASTM F1841-97 Standard Practice for Assessment of Hemolysis in Continuous Flow Blood Pumps
Standard operating procedure for assessment of hemolysis in continuous flow blood pumps



Copyright ©2024 All Rights Reserved