1.1 This guide is intended for use with controlled temperature comparison baths that contain test fluids and operate within the temperature range of –100 °C to 550 °C.
1.2 This guide describes the essential features of controlled temperature fluid baths used for the purpose of thermometer calibration by the comparison method.
1.3 This guide does not address the details on the design and construction of controlled-temperature fluid baths.
1.4 This guide describes a method to define the working space of a bath and evaluate the temperature variations within this space. Ideally, the working space will be as close as possible to isothermal.
1.5 This guide does not address fixed point baths, ice point baths, or vapor baths.
1.6 This guide does not address fluidized powder baths.
1.7 This guide does not address baths that are programmed to change temperature.
1.8 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.
1.9 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.
1.10 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.
ASTM E2488-22 Referenced Document
ASTM E1 Standard Specification for ASTM Thermometers
ASTM E344 Terminology Relating to Thermometry and Hydrometry*, 2023-12-01 Update
ASTM E644 Standard Test Methods for Testing Industrial Resistance Thermometers
ASTM E839 Standard Test Methods for Sheathed Thermocouples and Sheathed Thermocouple Material
ASTM E2488-22 history
2022ASTM E2488-22 Standard Guide for the Preparation and Evaluation of Liquid Baths Used for Temperature Calibration by Comparison
2009ASTM E2488-09(2014) Standard Guide for the Preparation and Evaluation of Liquid Baths Used for Temperature Calibration by Comparison
2009ASTM E2488-09 Standard Guide for the Preparation and Evaluation of Liquid Baths Used for Temperature Calibration by Comparison