ASTM E3376-23
Standard Practice for Calibration and Usage of Germanium Detectors in Radiation Metrology for Reactor Dosimetry

Standard No.
ASTM E3376-23
Release Date
2023
Published By
American Society for Testing and Materials (ASTM)
Latest
ASTM E3376-23
Scope
1.1 This standard establishes techniques for calibration, usage, and performance testing of germanium detectors for the measurement of gamma-ray emission rates of radionuclides in radiation metrology for reactor dosimetry. The practice is applicable only to samples of small size, approximating to point sources. It covers the energy and full-energy peak efficiency calibration as well as the determination of gamma-ray energies in the 0.06 MeV to 2 MeV energy region and is designed to yield gamma-ray emission rates with an uncertainty of ±3 % (see Note 1 ). This technique applies to measurements that do not involve overlapping peaks, and in which peak-to-continuum considerations are not important. Note 1: Uncertainty U is given at the 68 % confidence level; that is, where δ i are the estimated maximum systematic uncertainties, and σ i are the random uncertainties at the 68 % confidence level. Other techniques of error analysis are in use ( 1 , 2 ) . 2 1.2 Additional information on the setup, calibration, and quality control for radiometric detectors and measurements is given in IEEE/ANSI N42.14 and in Guide C1402 and Practice D7282 . 1.3 The values stated in SI units are generally to be regarded as standard. The rad is an exception. 1.4 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use. 1.5 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

ASTM E3376-23 Referenced Document

  • ASTM C1402 Standard Guide for High-Resolution Gamma-Ray Spectrometry of Soil Samples
  • ASTM D7282 Standard Practice for Setup, Calibration, and Quality Control of Instruments Used for Radioactivity Measurements
  • ASTM E170 Standard Terminology Relating to Radiation Measurements and Dosimetry

ASTM E3376-23 history

  • 2023 ASTM E3376-23 Standard Practice for Calibration and Usage of Germanium Detectors in Radiation Metrology for Reactor Dosimetry
Standard Practice for Calibration and Usage of Germanium Detectors in Radiation Metrology  for Reactor Dosimetry



Copyright ©2024 All Rights Reserved