BS 6955-12:1994
Calibration of vibration and shock pick-ups. Method of test for transverse shock sensitivity

Standard No.
BS 6955-12:1994
Release Date
1994
Published By
British Standards Institution (BSI)
Latest
BS 6955-12:1994
Replace
87/72304 DC-1987
Scope
ISO 5347 comprises a series of documents dealing with methods for the calibration of vibration and shock pick-ups. This part of ISO 5347 lays down detailed specifi-cations for the instrumentation and procedure to be used for transverse shock sensitivity testing. It applies to rectilinear accelerometers of the strain gauge, piezoresistive and piezoelectric type. This part of ISO 5347 is applicable for a time range from 0,01 ms to 10 ms and a dynamic range from 100 m/s to 10 m/s. The limits of uncertainty applicable are ± 20 % of reading.

BS 6955-12:1994 history

  • 1994 BS 6955-12:1994 Calibration of vibration and shock pick-ups. Method of test for transverse shock sensitivity