1 Scope
This document specifies the requirements and methods for the clinical
investigation of continuous automated non-invasive
sphygmomanometers used for the measurement of the
blood pressure of a patient.
This document does not cover usability aspects such as the form and manner of the
data display or output. This document does not specify a numerical threshold on the
minimum output
period . A continuous automated non-invasive
sphygmomanometer providing blood
pressure parameters (e.g., systolic blood pressure ,
diastolic blood pressure or mean arterial
pressure ) with an output period considerably
larger than 30 s is not typically considered a continuous automated non-invasive
sphygmomanometer .
This document covers both trending continuous automated non-invasive
sphygmomanometers and absolute accuracy
continuous
automated non-invasive sphygmomanometers and
focuses solely on requirements for the clinical investigation .
Representation of output is not covered by this document.
NOTE 1
IEC 62366‑1
provides requirements on the application of usability engineering to medical
devices. The usability engineering process can be used to
clarify for the intended user whether the displayed data concerns absolu...
BS EN ISO 81060-3:2022 Referenced Document
IEC 60601-2-34:2011 Medical electrical equipment - Part 2-34: Particular requirements for the basic safety and essential performance of invasive blood pressure monitoring equipment
IEC 80601-2-30:2018 Medical electrical equipment - Part 2-30: Particular requirements for basic safety and essential performance of automated non-invasive sphygmomanometers
ISO 14155:2020 Clinical investigation of medical devices for human subjects — Good clinical practice
ISO 14971:2019 Medical devices — Application of risk management to medical devices
ISO 81060-1:2007 Non-invasive sphygmomanometers - Part 1: Requirements and test methods for non-automated measurement type
BS EN ISO 81060-3:2022 history
2023BS EN ISO 81060-3:2022 Non-invasive sphygmomanometers - Clinical investigation of continuous automated measurement type