ISO 24384:2024
Water quality

Standard No.
ISO 24384:2024
Release Date
2024
Published By
International Organization for Standardization (ISO)  IX  /  ISO
Latest
ISO 24384:2024
 

Introduction

Analysis of the core content of the standard

ISO 24384:2024 has established for the first time an international standard method for the determination of chromium forms in water using LC-ICP-MS coupling technology. Its technological breakthroughs are reflected in:

Technical parametersPDCA chelation methodEDTA chelation method
Pretreatment temperature80±3℃70±3℃
Reaction time30 minutes60 minutes
Detection limit (52Cr)0.012-0.044μg/L0.022-0.030μg/L
Applicable water sample typesDrinking water/surface water/wastewater (0.2-500μg/L)

Method principle and technical evolution

The standard uses 2,6-pyridinedicarboxylic acid (PDCA) or ethylenediaminetetraacetic acid (EDTA) chelation pretreatment to uniformly convert unstable Cr(III) forms in water samples into stable complexes:

  1. pH is accurately adjusted to 6.9±0.1
  2. Forming Cr(III)-PDCA/EDTA complex
  3. LC separation followed by ICP-MS detection (monitoring 52Cr and 53Cr isotopes)

Compared with conventional spectrophotometry, this method reduces the detection limit of Cr(VI)/Cr(III) by more than 10 times and avoids errors caused by redox conversion.


Control of key interference factors

Typical interference case:

When >100mg/L transition metals (such as Co²⁺, Ni²⁺) are present in water samples, they will compete with chelating agents, resulting in low Cr(III) determination values. The standard recommends:

  • Sample dilution (maximum dilution ratio 1:10)
  • Spike recovery verification (required 80-120%)
  • Total chromium calibration (difference ≤30%)

Implementation recommendations

Based on the verification data of laboratories in 11 countries (Table C1/C2), it is recommended to give priority to the PDCA chelation scheme:

IndicatorsPDCA methodEDTA method
Cr(VI) recovery96-102%88-96%
Inter-laboratory deviation≤7.3%≤20%

Operation points: Use PEEK connecting tubes (inner diameter <0.25mm), verify the calibration curve daily, and store samples in a dark and refrigerated place (3±2℃).

ISO 24384:2024 Referenced Document

  • ISO 8466-1 Water quality — Calibration and evaluation of analytical methods — Part 1: Linear calibration function

ISO 24384:2024 history

Water quality

Standard and Specification




Copyright ©2026 All Rights Reserved
Update: Mon, 09 Mar 2026 08:31:54 +0000