This standard systematically specifies the grayscale calibration system of medical imaging display equipment for the first time, and establishes the grayscale standard display function (GSDF) based on the Barten human eye vision model to achieve accurate mapping of input level and output brightness/optical density. The main innovations include:
1. Dual media coverage: Applicable to both electronic display media (display/projector) and printed media (film/reflective sheet)| Dimensions | Electronic display media | Printed display media |
|---|---|---|
| Measurement parameters | Brightness (cd/m²) | Optical Density (OD) |
| Core Formula | Lj+1=Lj(1+Sj)/(1-Sj) | D(φ)=-log10[(L[Dm(φ)]-La)/L0] |
| Standard Function | Appendix B Brightness Lookup Table | Appendix C/D Density Lookup Table |
| Equipment Requirements | First-Class Brightness Meter (JJG 211) | Standard densitometer (JJG 452) |
The model integrates 8 physiological characteristic parameters of human eye:
The contrast sensitivity S(L) calculated by formula A.7 ensures that each grayscale step corresponds to 1 JND unit.
Synthetic uncertainty model (Formula 13):
u(Doutput)=√[u(Lm)²+u(LD)²]
Where u(LD) is calculated by the RMS of the brightness difference of 255 calibration points (Formula 14), the typical value should be <0.0015.