GB/T 43596-2023
The calibration method of the grayscale standard display function for image (English Version)

Standard No.
GB/T 43596-2023
Language
Chinese, Available in English version
Release Date
2023
Published By
General Administration of Quality Supervision, Inspection and Quarantine of the People‘s Republic of China  CN  /  GB
Latest
GB/T 43596-2023
 

Introduction

Analysis of the core content of the standard

This standard systematically specifies the grayscale calibration system of medical imaging display equipment for the first time, and establishes the grayscale standard display function (GSDF) based on the Barten human eye vision model to achieve accurate mapping of input level and output brightness/optical density. The main innovations include:

1. Dual media coverage: Applicable to both electronic display media (display/projector) and printed media (film/reflective sheet)
2. Physiological adaptation: Ensure that the grayscale steps meet the just noticeable difference (JND) of the human eye through the visual model in Appendix A
3. Dynamic adaptation: Supports device calibration in the luminance range of 10-4~103 cd/m²

Comparison of calibration technology frameworks

Dimensions Electronic display media Printed display media
Measurement parameters Brightness (cd/m²) Optical Density (OD)
Core Formula Lj+1=Lj(1+Sj)/(1-Sj) D(φ)=-log10[(L[Dm(φ)]-La)/L0]
Standard Function Appendix B Brightness Lookup Table Appendix C/D Density Lookup Table
Equipment Requirements First-Class Brightness Meter (JJG 211) Standard densitometer (JJG 452)

Key technologies of Barten vision model

The model integrates 8 physiological characteristic parameters of human eye:

  1. Nervous system noise threshold
  2. Retinal photoreceptor cell discreteness
  3. Pupil diameter dynamic adjustment (Formula A.6)
  4. Photon quantum efficiency (357×3600 photos/td sec deg²)
  5. 0.1s time integration characteristics
  6. Spatial frequency response (0.2-50 c/deg)
  7. Optical modulation transfer function (Formula A.1)
  8. Directional sensitivity difference

The contrast sensitivity S(L) calculated by formula A.7 ensures that each grayscale step corresponds to 1 JND unit.


Implementation Suggestions

Electronic Display Calibration Process:
  • Under <0.05 cd/m² darkroom environment
  • Use a 2° field of view luminance meter to measure 256 levels of grayscale
  • Generate LUT by interpolation according to Appendix B
  • Verify u(LD)<0.0015
Medical Film
Calibration Key Points:
  • The transmission type needs to control the L0=2000 cd/m² viewing light
  • The reflection type standard observation surface brightness is 1300 cd/m²
  • The density meter needs to meet the error of ±0.02 when D≤2.0

Uncertainty Analysis

Synthetic uncertainty model (Formula 13):

u(Doutput)=√[u(Lm)²+u(LD)²]

Where u(LD) is calculated by the RMS of the brightness difference of 255 calibration points (Formula 14), the typical value should be <0.0015.

GB/T 43596-2023 Referenced Document

GB/T 43596-2023 history

  • 2023 GB/T 43596-2023 The calibration method of the grayscale standard display function for image
The calibration method of the grayscale standard display function for image

Topics on standards and specifications

Standard and Specification

CNS 14176-14-2007 Digital imaging and communications in medicine (DICOM) - Part 14: Grayscale standard display function DIN EN IEC 62563-2:2020-12 Medical electrical equipment - Medical image display systems - Part 2: Acceptance and constancy tests (IEC/CDV 62563-2:2020); German and English version prEN IEC 62563-2:2020 / Note: Date of issue 2020-11-20 DIN EN IEC 62563-2:2020 Medical electrical equipment - Medical image display systems - Part 2: Acceptance and constancy tests (IEC/CDV 62563-2:2020); German and English version prEN IEC 62563-2:2020 DIN EN 62563-1/A1:2016-05 Medical electrical equipment - Medical image display systems - Part 1: Evaluation methods (IEC 62B/983/CDV:2015); German and English version EN 62563-1:2010/FprA1:2015 / Note: Date of issue 2016-04-01*Intended as an amendment to DIN EN 62563-1 (2014-01). DIN EN 62563-1/A1:2016 Medical electrical equipment - Medical image display systems - Part 1: Evaluation methods (IEC 62B/983/CDV:2015); German and English version EN 62563-1:2010/FprA1:2015 UNE-EN 62563-1:2010 Medical electrical equipment - Medical image display systems -- Part 1: Evaluation methods (Endorsed by AENOR in June of 2010.) CNS 14275-1998 Digital compression and coding of continuous-tone still images Part 1: Requirements and quidelines DS/EN 61223-2-5:2013 Evaluation and routine testing in medical imaging departments - Part 2-5: Constancy tests - Image display devices SMPTE RP 2080-2:2014 RP 2080-2:2014 - SMPTE Standard - Measurement and Calibration Procedure for HDTV Display Luminance Levels and Chromaticity RP 2080-2:2014 RP 2080-2:2014 - SMPTE Standard - Measurement and Calibration Procedure for HDTV Display Luminance Levels and Chromaticity