GB 5009.241-2017 integrates and replaces six historical standards, and systematizes the magnesium determination methods that were previously scattered in various food testing standards. The most significant technological advancement of the new standard is reflected in:
| Comparison dimensions | 2003 version of the standard | 2017 version of the standard |
|---|---|---|
| Method system | Flame atomic absorption spectrometry only | Three methods in parallel (ICP-OES/ICP-MS newly added) |
| Pretreatment technology | Traditional wet digestion as the main method | Four methods optional (microwave/pressure digestion, etc.) |
| Scope of application | Specific food categories | Covering all food categories |
Use atomic absorption spectrometer at a wavelength of 285.2nm. Key operating points:
Use the detection method of GB 5009.268, suitable for high-throughput detection of large quantities of samples:
| Parameters | ICP-OES | ICP-MS |
|---|---|---|
| Detection Limit | 0.05mg/kg | 0.001mg/kg |
| Isotope | - | 24Mg/25Mg/26Mg |
During the inspection of a dairy company, it was found that when using microwave digestion (power 1600W) to treat milk powder samples, the recovery rate was increased by 12% compared with traditional wet digestion, and the detection time was shortened by 60%.