This part of the standard specifies the method principle for determining the activity of alkaline phosphatase (ALP) in cheese using continuous fluorescence kinetic method: using a non-fluorescent aromatic monophosphate substrate 2′-[2-benzothiazolyl]-6′-hydroxybenzothiazole phosphate, it is hydrolyzed under the action of ALP to produce a highly fluorescent product, and the fluorescence is detected at 38°C for 3 minutes using a Fluorophos® instrument.
| Version | BS EN ISO 11816-2:2016 | BS EN ISO 11816-2:2024 |
|---|---|---|
| Instrument Requirements | FLM200 Model | Upgrade to FLM300 (Integrated Heating Block) |
| Calibration Procedure | Step-by-Step | Optimization of User Interface and Electronics |
| Detection Limit | 7,000 mU/kg | Maintain the same sensitivity |
Special processing requirements for different types of cheese:
Use three calibration solutions (A:0μmol/L, B:17.24×10⁻³μmol/L,
| Control Type | Qualified Standard | Frequency |
|---|---|---|
| A/D Test | 302 FLU±4 FLU | Daily Startup |
| Instrument Control Solution | 602 FLU±12 FLU | Daily Testing |
| Negative control | <10 mU/kg | Each batch |
ALP activity calculation formula:
Ap = (Ap1 × f2 × 25)/m
Wherein:
Ap1=supernatant ALP activity (mU/L), f2=dilution factor, m=sample mass (g)
Verification steps when the test value is abnormal:
This method is not suitable for cheeses containing mold inside (such as blue cheese), and special attention should be paid to the sampling location for large hard cheeses that have been blanched at more than 50°C.