The National Standard of the People's Republic of China GB/T 43530-2023 "Performance Test Method of Lobster Eye Focusing Optical Components" is under the jurisdiction of the National Technical Committee for Photoelectric Measurement Standardization and jointly drafted by the National Astronomical Observatory of the Chinese Academy of Sciences and other units. This standard mainly regulates the performance test method of lobster eye focusing optical components (MPO) used in space astronomy, medical imaging, customs inspection and other fields.
| Standard dimensions | GB/T 43530—2023 | Comparative analysis |
|---|---|---|
| Scope of application | X-ray focused imaging fields such as space astronomy, medical imaging, and customs inspection | Added application specifications in the medical imaging field |
| Core parameters | X-ray focal length, angular resolution, effective area, etc. | Introduction of half-energy width indicators to enrich test dimensions |
| Test conditions | Standard atmospheric conditions (temperature and humidity), vacuum environment requirements | New cleanliness requirements (level 8 and above) |
In the field of space astronomy, the performance of MPO directly affects the sensitivity and resolution of the X-ray imaging system. The implementation of GB/T 43530-2023 standard can ensure that MPO components produced by different manufacturers have unified technical indicators, thereby improving the overall performance of space observation equipment.
The MPO performance test device consists of an X-ray detector, a theodolite, a half-reflecting half-mirror and an X-ray light source. Its working principle is based on grazing incidence reflection, and the quantitative evaluation of the performance of the MPO component is achieved through the change in the shape of the converging spot.
Key parameters: X-ray source voltage ≤ 20kV, focal spot diameter ≤ 0.05mm, detector single pixel size ≥ 0.04mm.It is recommended to conduct the test in an environment with a cleanliness level of 8 or above, using the clean room design specifications of the GB50073 standard. The equipment parameters must be calibrated before testing, and all environmental factors must be recorded.
The samples should be stored in a dry and clean environment, and it is recommended to be filled with protective gas (such as nitrogen). During the test, use special fixtures to handle and place with care to avoid the influence of dust and salt spray.
Fill in the test report record form in strict accordance with the format of Appendix A of the standard. Pay special attention to the calculation error control of the PSF fitting parameters. It is recommended to use the least squares method to optimize the fitting accuracy.