Technical indicators | Required values | Test methods | Clinical significance |
---|---|---|---|
Mass resolution | <0.8u(Li/In/Tl) | 10% peak width method | Ensure the mass spectrum peak separation between elements |
Detection sensitivity | Hg≥1×106cps/(mg/L) | Signal-to-noise ratio method | Ensure the detection capability of trace elements |
Oxide Interference | <3.0%(CeO+/Ce+) | Standard Solution Method | Control Polyatomic Ion Interference |
Take the carryover contamination rate test of Inductively Coupled Plasma Mass Spectrometer as an example: high/low concentration alternating injection mode is required, calculated by the formula Ki=(Ai4-Ai6)/(Ai3-Ai6)×100%, and the requirement is ≤0.05%. This indicator directly affects the accuracy of continuous detection of clinical samples.
Compared with the GB/T 34826-2017 general ICP-MS standard, this document adds:
1. Specific requirements for whole blood matrix samples (component proportion > 85%)
2. Supporting detection methods for clinical element groups (12 items including K/Ca/Fe/Se)
3. EMC/safety requirements specific to medical devices (referenced to standards such as YY0648)