| Comparison dimensions | 2008 version | 2019 version |
|---|---|---|
| Structural framework | Management requirements + technical requirements | Five parts: general/structure/resources/process/management system |
| Risk management | No special chapter | Added 8.5 Measures to address risks and opportunities |
| Judgment rules | Not clearly defined | Added 3.7 Terminology and 7.1.3 Implementation requirements |
The new version of the standard requires laboratories to establish an impartiality risk identification mechanism, with special emphasis on the assessment of potential impacts such as ownership relationships and financial transactions. Implementation case: A third-party testing agency established a commercial conflict of interest assessment form and updated the list of related parties every quarter.
Specify the equipment calibration status identification requirements (6.4.8), and recommend the use of a three-color label system: green (valid within the period), yellow (about to expire), red (disabled). Key equipment such as electronic balances need to implement period verification procedures.
11 new monitoring methods, including the following:
Requires the application of the judgment rules to be clarified, for example:
When the measured value ±U≤limit value (where U is the expanded uncertainty), it is judged to be in complianceThe basis for the judgment rules used must be stated in the report.
It is recommended to establish a two-tier document structure:
| First-level document | Policy manual | Covering 8.2 management system requirements |
| Second-level document | Process flow chart | Visualize 7.1-7.11 processes |